Non-invasive blood pressure accuracy study for adults
Official title: Non-Invasive Blood Pressure Accuracy Study for Adult Subject Populations
Run by GE Healthcare · NCT07628296 · Data updated 2026-11-12
This study tests whether non-invasive blood pressure readings match invasive aortic pressure measurements in adults who need arterial line monitoring.
Quick facts
| Study type | Interventional |
|---|---|
| Ages | 18 Years and up |
| Sex | All |
| Enrollment | 130 (estimated) |
| Sponsor | GE Healthcare Industry-sponsored |
| Locations | 1 site (Tampere) |
| Status | Recruiting |
| Started | May 19, 2026 |
| Main results expected | December 31, 2026 |
| Study ends | June 30, 2027 |
| Registry record updated | July 8, 2026 |
| Trial ID | NCT07628296 on ClinicalTrials.gov |
Is this trial a fit for you?
✓Likely a good fit
Adults aged 18 and over with appropriate cuff size and an arterial line as part of standard care, who can consent and stay in the unit for the procedure.
✕Probably not a fit
Patients without an arterial line or cuff-fit limitations may not benefit from this study's measurements for clinical decision-making.
A plain-English reading of the study's rules. Only the study team can confirm whether you qualify.
What this trial studies
This interventional study collects paired non-invasive blood pressure (NIBP) and invasive aortic pressure data in a clinical setting to validate NIBP measurement algorithms. It follows IEC 80601-2-30 and ISO 81060-2 standards for accuracy of automated NIBP devices. Eligible participants are adults with a suitable cuff size and an arterial line as part of standard care, and who can consent and stay in the unit for the procedure. The study uses the NIBP Standard Parameter Module to assess algorithm accuracy across real-world conditions. Data from the study will inform device performance and potential improvements in NIBP measurement systems.
Why it matters
Potential benefit: If successful, the study could improve the accuracy of automated NIBP devices, potentially leading to better blood pressure monitoring in clinical care.
How similar studies have gone: Past work under similar IEC and ISO references has demonstrated measurement alignment between non-invasive and invasive pressures in controlled conditions, with ongoing efforts to validate automated devices.
Who can join: the full rules
Show full inclusion / exclusion criteria
Inclusion Criteria: 1. Age requirement: 18 and over. 2. Have an upper limb (right OR left side) that fits the cuff size of the device (circumference ranging 17 cm to 40 cm) OR have a thigh (right OR left side) that fits a cuff size of the device (circumference ranging 38 to 50 cm). 3. Have an arterial line as a standard of care, which measures blood pressure at the aortic root. 4. Able to provide written informed consent. 5. Expected to remain in the unit for the duration of the study procedure. Exclusion Criteria: 1. Have previously participated in this study (no subject may participate more than once). 2. Pregnant and lactating women per site standard of care. 3. Subjects with peripheral arterial diseases (causing left/right arm difference) 4. Hemodynamically unstable subjects at the discretion of the investigator. 5. Exhibited injuries, deformities, intravenous lines, or other abnormalities that, in the opinion of the Investigator, may prevent proper cuff application or functioning. 6. Are suffering from infection(s) or immunocompromised patients that require isolation. 7. PI or designee decision due to subject health conditions
Where it's running
-
Tampere Heart Hospital
Tampere, Finland Recruiting
Similar studies
Other studies for the same condition or treatment. Recruiting studies are listed first.
-
Automated Non-Invasive Blood Pressure: Tracking Blood Pressure Changes Using SpO2 Data
This study tests an algorithm that estimates blood pressure changes using data from an SpO2 sensor in patients in the OR, PACU, or ICU who already have invasive BP measured as part of standard care.
-
NIBP Accuracy Validation in Adults
This study tests how accurately a non-invasive blood pressure device measures blood pressure in adults by comparing it with arterial line readings.
-
New Non-Invasive Blood Pressure (NIBP) Enhancement for Neonates and Children
This study tests a new NIBP device enhancement in neonates and kids to see if it can measure blood pressure more accurately.
-
Inflation-based non-invasive blood pressure test compared with standard cuff method
This study tests whether the Philips iNIBP device gives BP readings that match the usual auscultatory cuff method in people aged 3 to 85.
-
Overnight Temperature Regulation to Test Gut Health, Inflammation, and Circadian Rhythms
This study tests whether two weeks of using the Eight Sleep Pod at night can improve gut bacteria balance, lower inflammation, and strengthen daily temperature and blood pressure rhythms in postmenopausal women and age-matched men with poor sleep.
-
Blood Pressure from Face Video (Vitacam)
A contactless vitals tool called Vitacam tests if video-based blood pressure readings from the face can match standard cuff measurements in adults.
How to take part
- Review the eligibility criteria above with your treating physician.
- Check the official listing on ClinicalTrials.gov for the latest contact details and recruitment status.
- Contact the study coordinator or principal investigator to ask for pre-screening. Pre-screening is free and never obligates you to enroll.